Refrigerators Freezers Spark-Proof |
|
Spark-Proof Refrigerators and Freezers (+4°C Refrigerator / -18°C Freezer)
|
|
|
|
|
Refrigerator Model TR-240-SP
|
Freezer Model (-18ºC) TF-180-SP
|
The model TRF-SP stand suits all spark
proof refrigerator and freezer models
offered to raise them to a minimum of
300mm above the floor as required.
|
NOTE: Only the interior workspace is spark-proof
|
<>
Features
These spark proof refrigerator and freezer models have been specifically designed to allow for the safe storage of volatile samples which are unsafe for storage in standard refrigerators and freezers. All cabinets are completely spark free on the inside. This ensures that no electrical components, which may cause a spark in the event of a malfunction, are fitted within the workspace. These models meet the Ex'n non-sparking equipment requirments and for added safety, the refrigeration compressor is fitted with a unique ‘spark proof start mechanism' should any flammable vapours be present outside the refrigerator at floor level.
In reference to AS/NZS60079.10:2004 for zoning of hazardous areas:
For zone 2 classification it is recommended that a refrigerator or freezer be located 300mm above floor level. The model TRF-SP stand suits all spark proof refrigerator and freezer models offered to raise them to a minimum of 300mm above the floor as required.
| (TR) Spark-Proof Refrigerator +2°C to +6°C |
(TF) Spark-Proof Freezer -15°C to -18°C |
• All electrical parts and controls are outside working space
• "Wet Wall" evaporator positioned behind the rear liner
• Natural gravity convection air movement
• Solid state electronic display
• Plastic inbuilt moulded liner (PS plastic has good antistatic
properties)
• Powdercoated shelves and transparent storage drawers
• Inner condensate drains through a water trap to an external
auto disposal container.
• Floor standing
• Cyclic Defrost
• Quiet operation
• Solid access door with magnetic gasket
• Microprocessor controller with built in Error codes
• Option al alarm contacts available |
• All electrical parts and controls are outside working
space
• Natural gravity convection air movement
• Manual defrost
• Solid state electronic display
• Plastic inbuilt moulded liner
• Freezing coils positioned above each draw/bin
compartment
• Fixed position pull out PVC drawers/bins
• Floor standing
• Quiet operation
• Solid access door with magnetic gasket
system.
• Microprocessor controller with built in Error codes
• Optional alarm contacts availableble |
MODEL
|
TR-130-SP
|
TR-240-SP
|
TR-270-SP
|
TF-90-SP
|
TF-180-SP |
| Capacity |
130 Litres
|
240 Litres
|
270 Litres
|
90 Litres
|
180 Litres
|
Internal Dims
|
70x46x43 |
127x45x43 |
146x46x43
|
65/43x40x40 |
124x40x38 |
External Dims
|
84x54x53 |
151x55x56 |
165x53x57
|
87x54x60 |
151x55x56 |
Temperature Range
|
+2°C to +6°C |
-15°C to -20°C |
| Shelves |
3 |
5 |
5 with 1 drawer |
3 |
4 |
| Electrical |
1A 240V |
Defrost Method
|
Cyclic |
Cyclic |
Cyclic
|
Maunal |
Manual
|
| Weight |
50 |
66 |
99
|
45 |
66 |
Click to download PDF Product Sheet
|